연동 펌프 SR400 그것은 높은 유량의 SN 시리즈 및 YZ 시리즈 펌프 헤드와 함께 사용할 수 있습니다, LCD 화면은 사용자가 속도를 유연하게 제어 할 수 있도록 현재 펌핑 속도를 보여줍니다, 풋 페달은 침투 지방 흡입 펌프로 펌프를 사용할 때 선택 사항이며, 많은 분야에서 액체 이송을위한 높은 비용 성능 펌프입니다.
Model No. | SR400-YZ1515X/YZ2515X/SN15/SN25 |
Pump Roller | SST 3/6 rollers |
Pump Head | ABS engineering plastic (YZ) PC engineering plastic (SN) |
Fast Tubing | YES |
Pump Color | White/Black (YZ) Transparent (SN) |
Motor Type | Stepper motor |
Motor Speed | 0-400rpm |
Flow Range | 0-1700ml/min |
Tubing Size | ID1.6-8.0mm WT1.6-2.5mm OD≤11.4mm |
Tubing Life | Silicone≥200H BPT Rubber≥1000H |
Pressure Rating | Up to 2 bar(30psi) |
Working Environment | 0-40℃,<80% non-condensing |
Max DB | 65db |
Suction | 5m |
Head | 8m |
Dimension (L*W*H) | 301*151.5*233mm |
Net Weight | 5508g(3 roller)/5608g(6 roller) |
Picture | Material | Features | Service Life |
Silicone | Food grade, High flexibility Temp. range -4℃ - +180℃ For non-corrosive liquid | ≥200h | |
Pharmed BPT | Saint-Gobain, FDA standard Temp. range -51℃- +132℃ Anti weak acid and alkali | ≥1000h | |
Viton | MasterFlex, FDA Standard Temp. range -20℃- +260℃ Anti strong acid and alkali | ≥1000h |
1. Tubing life measured by pumping pure water under continuous working at the room temperature 20℃ and no extra pressure until it cracks.
2. Tubing life vary from different medium, motor speed, working environment, the lower motor speed helps to extend the service life of tubing.
3. Please replace new pump tube regularly to achieve higher precision liquid transfer.
Flow Rate Instruction (YZ1515X)
Tubing | Speed (rpm) | Tubing size (ID*WT: mm) | 3 Roller Flow rate (ml/min) | 6 Roller Flow rate (ml/min) |
Rubber tube | 0-400rpm | 1.6*1.6 | 0-113 | 0-88 |
3.2*1.6 | 0-462 | 0-347 | ||
4.8*1.6 | 0-966 | 0-612 | ||
6.4*1.6 | 0-1447 | 0-804 | ||
8.0*1.6 | 0-1887 | 0-1332 | ||
Silicon tube | 1.6*1.6 | 0-117 | 0-92 | |
3.2*1.6 | 0-491 | 0-375 | ||
4.8*1.6 | 0-1020 | 0-652 | ||
6.4*1.6 | 0-1494 | 0-848 | ||
8.0*1.6 | 0-1700 | 0-1400 |
Function of Foot Pedal
Step 1 Connect the plug into power supply AC110-240V
Step 2. Connect the foot pedal into the interface at back of the pump
Step 3. Press the power (main) switch at back of the pump
Step 4. Press the red ON/OFF button at front of the pump to start the pump, at this time you can see shaft of pump is turning
Step 5. You can gently turn the SPEED button at front of the pump to adjust the speed
Remark 1: The pump must be started working under lower speed,so SPEED button must stay at nearly middle (or less at CCW direction) position before start working
Remark 2: Foot pedal only works when the red button at front of the pump staying ON (Foot pedal will NOT works when the red button at front of the pump staying OFF)
Remark 3: Foot pedal must be connected with the pump, or the buttons on the pump will lose its functions
Remark 4: SPEED adjustment only works when pump is working (shaft of pump is turning)(SPEED will not be adjustable when pump is stopping)
Dimension (unit: mm)